PGC EnCal 3000
L’ EnCal 3000 è un gascromatografo progettato specialmente per la misura fiscale del gas naturale. Usa componenti cromatografici basati sulla più recente tecnologia dei sistemi microelettromeccanici (MEMS). Questo permette di avere ottimi risultati di ripetibilità e di analisi accurata. L’analizzatore compatto contiene l’hardware analitico, selezione del canale di misura e tutta l’ elettronica necessaria per lavorare in autonomia.
| Dati tecnici | Downloads | ||||
EnCal 3000: Energy measurement analyser
| Analytical hardware |
2 parallel isothermal GC modules with narrow-bore capillary column technology in combination with MEMS based analytical components |
| Analysis output |
Full composition of any natural gas up to C6+ or C9+ (option), Heating value, Density, Wobbe index |
| Component range |
N2: 0.005 - 20 % neo-C5: 0.005 - 0.25 % CH4: 65 - 100 % C5: 0.001 - 0.25 % CO2: 0.001 - 20 % C6: 0.001 - 0.10 % C2: 0.001 - 10 % C7: 0.001 - 0.05 % C3: 0.001 - 10 % C8: 0.001 - 0.05 % C4: 0.001 - 10 % C9+: 0.001 - 0.05 % Concentrations outside these ranges on request |
| Performance: | Uncertainty: < 0.1 % for all calculated properties Repeatability: < 0.01 % for all calculated properties Detection limit for C5: 1 ppm Analysis time: 3 min. for C6+ analysis, 5 min. for C9+ analysis |
| Ambient conditions |
Temperature: -20 °C to +55 °C |
| Dimensions | Base diameter 37 cm x Height 37 cm (diameter 14" x Height 14") |
| Weight | < 30 kg |
| Approvals | ATEX II2G E Ex d IIB T4 (IIC certification pending) NEC Class I, Div. I, Groups A, B, C and D (pending) IP 66, Vibration and shock test in acc. with IEC 60068-2-31 and 64 EMC according to EN 61000-6-2 and EN 61000-6-4 PTB-approval ID 7.614/06.56 |
| Power supply |
24 V DC, 18 W nominal (50 W start-up peak) for non-heated version 24 V DC, 120 W nominal (170 W start-up peak) for heated version (ambient < 0 °C) |
| Interfaces |
Ethernet UTP 10 Base-T for ModBus TCP/IP and PC link |
| Analyser | Complete stand-alone operation, including all calculations and generation of report formats, without need for operator intervention. Calculations in acc. with ISO 6976, GPA 2172 or GOST 22667 |
| PC requirements | Windows 2000 or Windows XP professional edition (Service pack 1 or higher) 1000 MHz processor, 512 MB RAM, CD-ROM player, free Ethernet port |
| Data logging | History Log: local storage of last 35 days of all analytical data (analysis, events, alarms, averages, last chromatogram, calibration data) in accordance with API Report 21.1. All data available on remote workstation in XML format |
| Sample conditioning |
Integral part of analyser. Consists of pressure regulators for each stream, moisture trap, particle filters and double block-and-bleed stream selection for up to 5 streams and 1 calibration gas |
| Helium | Quality N45 or better, supply pressure 5.5 barg, consumption +/- 8 ml / min |
| Calibration gas | Supply pressure 1 barg nominal. Consumption +/- 600 ml / day @ atm. pressure Composition depending on application |